Description

Turbidity Point Analyzer 

The LS-A2180S Automatic Turbidity Point Analyzer employs modern high-tech microelectronic control technology, using a PLC as the system control core.

 

Standards

Applicable Standards: GB/T 6986, ASTM D2500

 

Feature

1. 7-inch intelligent touchscreen and perpetual calendar clock display.

2. Chinese LCD display, human-machine menu-driven dialogue, guided operation, fully automated turbidity point determination process.

3. The entire instrument has a reasonable structure, is simple to operate, and is safe and reliable.

Turbidity Point Analyzer 

Technical Parameters

• Applicable oils  Light oils with a cloud point range of 10℃ to -65℃;
• Measurement repeatability  2℃
• Display resolution  0.1℃; Accuracy: 0.3%
• Cooling limit depth  <=-70℃ (dual-compressor cascade refrigeration)
• Cooling rate  Temperature difference >15℃ within 10 minutes
• Sample volume  Cloud point: 45ml per sample;
• Number of samples tested  2
• Cold bath  2-well (aluminum bath)
• Rated power  1500W
• Power supply voltage  220V±10% 50H
• Ambient temperature  5℃~40℃
• Relative humidity  <85%
• Dimensions  400mm×790mm×490mm
• Instrument weight  50kg