Description

Automatic acid value titrator

The LS-A1045 Automatic Acid Value Analyzer employs the potentiometric titration principle. By recording the electrode potentials and titration volume during the titration process, it identifies the equivalence point and the corresponding volume of the standard titrant, thereby determining the acid or alkali value of the sample. The instrument can accurately detect the acid value of petroleum products such as transformer oil, turbine oil, fire-resistant oil, diesel, and gasoline. It is applied in industries such as chemical, power, petroleum, environmental protection, and railway.

 

Standards

Compatible Standards: GB/T 7304, GB/T 18609, ASTM D664

 

Feature

1. Windows operating platform, extra-large screen, touch operation, and workstation functionality.

2. Real-time display of titration curves; storage and printing of titration curves, results, and data.

3. Uses imported titration units, resulting in higher measurement accuracy and better stability.

4. Automatic cleaning and automatic constant-value liquid addition.

5. Automatic endpoint determination and automatic filtering of false endpoints; manual endpoint selection is also possible.

Automatic acid value titrator

Technical Parameters

• Measurement range  Greater than 0.01 mg KOH/g
• Accuracy  Relative error ≤ 5%
• Potential measurement range  -200~700 mV
• Basic error  0.1%FS ±0.5 mV
• Burette volume  10 ml
• Minimum burette volume  0.01 ml
• Burette accuracy  ±0.1%FS
• Dimensions  350 mm × 450 mm × 530 mm
• Instrument weight  13.6 kg
• Printer optional