Ozone aging test chamber
- Description
- Inquiry
Description
Ozone aging test chamber
I. Applications
Ozone, though present in very low atmospheric concentrations, is a major factor in rubber cracking. This product simulates and intensifies atmospheric ozone conditions to study the effects of ozone on rubber and provide a method for rapidly identifying and evaluating rubber’s resistance to ozone aging and the protective effectiveness of antiozonants.
Reference standards include: GB/T7762-2003, Vulcanized or Thermoplastic Rubber – Static Tensile Test Method for Ozone Cracking Resistance, GB/T2951.21-2008, General Test Methods for Cable Insulation and Sheathing Materials, GB/T 11206-2009, Rubber Aging Test – Surface Cracking Method, and test methods specified in other relevant standards.
II. Principle and Method
Test methods can be categorized as “static” or “dynamic,” depending on the test method and standard. Static testing involves stretching the sample before placing it in a test chamber for testing. Dynamic testing involves placing the sample in the test chamber’s fixture and testing while it is stretched. The tensile fluctuation should be approximately 5%-45% of the sample itself.
III. Structural Features
3.1. Structure: All-in-one unit.
3.2. Materials
Inner Box: SUS304 stainless steel
Outer Box: SUS304 stainless steel or painted
Insulation: Dual protection with rigid polyurethane foam and aluminum silicate insulation.
3.3. Door – Single door, opening on the left.
Two silicone rubber seals on the window and door frames, and an electric heater to prevent condensation.
Observation windows (W300 x H450 mm) and lighting are provided above the doors and windows.
3.4. Sample Holder Requirements
a) (Dynamic Test; optional) – Secure the specimen at both ends with clamps at the required elongation. When exposed to ozone-containing air, the specimen’s length should be approximately parallel to the airflow.
b) The clamps should be made of a material that does not readily decompose ozone (e.g., aluminum). c) Maintain a specimen rotation speed between (20 and 25) mm/s. Each specimen should continuously move along the same path within a plane perpendicular to the dry airflow.
d) The time required for one rotation of the specimen is (8 to 12) minutes. The specimen scanning area should be at least 40% of the effective area of the test chamber.
e) Specimen preparation should comply with the requirements of GB/T 9865.1. Specimens should preferably be cut from freshly molded test pieces; if necessary, they may be cut from finished products. A minimum of three specimens should be used for each test condition. Standard strip specimens should be at least 10 mm wide and 2.0 mm ± 0.2 mm thick. The specimen length between the two ends of the grips before tensioning should be at least 40 mm.
3.5. Control Panel – Controller Display
Power Switch
Light Button
3.6. Chamber Configuration – One internal stainless steel SUS #304 rack turntable
One φ50 mm wire hole, with a cover and silicone rubber plug.
4 movable casters
4 horizontal fixed legs
A drain hole is located at the lower rear of the unit to facilitate condensate drainage.
Ozone aging test chamber
Technical Parameters
| Temperature Range | 0°C-70°C |
| Temperature Fluctuation | ±2.0°C |
| Ozone Concentration Accuracy | 10-1000 pphm ±10% |
| Sample Holder Lifting Frequency | 2 R/min |
| Airflow Speed | 12-16 mm/s |
| Test Gas Flow Rate | 20-70 L/min |
| Sample Holder Rotation Speed | 360° (1-3 rpm) |
| Humidity Range | 20-98% |