Dynamic Thermomechanical Analyzer
Dynamic Thermomechanical Analyzer applies single-frequency or multi-frequency oscillatory forces to samples at a programmable temperature to study their mechanical behavior. The instrument measures deformation, storage modulus, loss modulus, and loss factor as a function of temperature, time, and force frequency. Suitable specimens include thermoplastics and thermosetting plastics, rubber, coatings, metals and alloys, inorganic materials, and composite materials.
- Description
- Inquiry
Description
Dynamic Thermomechanical Analyzer
Key Performance Indicators
The instrument applies single-frequency or multi-frequency oscillatory forces to samples at a programmable temperature to study their mechanical behavior. The instrument measures deformation, storage modulus, loss modulus, and loss factor as a function of temperature, time, and force frequency. Suitable specimens include thermoplastics and thermosetting plastics, rubber, coatings, metals and alloys, inorganic materials, and composite materials.
Dynamic Thermomechanical Analyzer
Technical Parameters
| Temperature Range | -150°C to 600°C |
| Instrument Structure | Circular, sealed furnace ensures uniform heating, with bottom-mounted sample loading. |
| Heating Rate | 0.1~ 20°C/min |
| Temperature Accuracy | ±0.1°C |
| Frequency Range | 0.01~100 Hz (Selectable) |
| Force Range
|
1mN~18N; Minimum 0.001N (dynamic)/0.005N (static), Maximum 50N, Resolution 0.0001N |
| Displacement Resolution | 1nm |
| Deformation Range | Maximum 25mm |
| Amplitude | ±0.5~10000µm |
| Modulus Range | 10³~10¹² Pa |
| Modulus Accuracy | 0.5% |
| Damping Range (tanδ): | 0.005~100 |
| Tanδ Resolution | 0.00005 |
|
Sample Temperature Control (STC) Functions, software screenshots provided. |
|
|
Deformation Modes: Compression, Penetration (maximum testable sample diameter no less than 15mm, maximum thickness no less than 25mm); Tensile (maximum width no less than 13mm, maximum thickness no less than 5mm, maximum length no less than 30mm); Shear, Three-Point Bending, Single/Dual Cantilever. |
|
|
Automatically identifies installed sample holders, automatically calibrates, and automatically measures sample length. |
|
|
The software features test editing, data acquisition, Storage, processing, and analysis functions; software controls purge and shielding gases. 1.Software Features Experimental Design Curing Simulation: Supports step-by-step temperature ramping and constant temperature testing, automatically recording modulus-time curves. Time-Temperature Equivalence (TTS): Generates master curves for long-term performance predictions. 2. Software Features Data Processing Automatic Calculation: Real-time output of storage modulus, loss factor (tanδ), and glass transition temperature (Tg). Multi-Graph Comparison: Supports overlay analysis of data at different frequencies and temperatures. |
|
|
Optional Special Accessories |
Humidity Accessory |