Use rotary viscometer to test starch viscosity
introduction
The basic principle of measuring starch viscosity with rotary viscometer is to use the gelatinization behavior of the sample in a certain temperature range and measure its viscosity change by rotating viscometer. As the temperature increases, the starch sample will gradually gelatinize, thus changing its viscosity. Rotary viscometers can accurately measure this change and provide quantitative data on viscosity values.
principle
The gelatinization process of starch means that under heating conditions, starch particles absorb water and expand and gradually break, releasing starch molecules to form a thick paste substance. With the increase of temperature, the water in starch particles increases, which weakens the interaction of starch molecular chains, causing the particles to expand and eventually collapse. In this process, the viscosity of starch will increase significantly with the increase of temperature. By monitoring the effect of temperature changes on starch gelatinization, detailed information on starch viscosity can be obtained.
A rotary viscometer determines the viscosity of a liquid by measuring the resistance of a rotating rotor. The rotor inside the instrument rotates in the sample liquid, and the viscosity of the liquid causes resistance to the rotation of the rotor, which is recorded by the sensor of the rotor and the instrument. The viscometer calculates the viscosity of the liquid by measuring the relationship between the torque of the rotor and the rotating speed. During the measurement process, the sample is kept at a constant temperature so that the influence of temperature on viscosity can be accurately reflected.
Test procedure
1. Sample mixing
1.1 Cold water is not paste starch treatment method
Weigh the appropriate starch sample (accurate to 0.1 g) and place it in a four-mouth flask.
Use the measuring cylinder to add distilled water or deionized water to ensure that the mass concentration of the dry base solid of the sample reaches the set value, and the total volume is about 2/3 of the flask volume.
Place the four flask in a constant temperature water bath that has been heated to 55°C.
Install agitator, condensing tube and thermometer, turn on condensate and agitator.
Heat the thermostatic water bath to raise the sample temperature to 92°C to 95°C.
After maintaining the temperature and stirring for a certain time, the sample was taken for viscosity measurement.
1.2 Starch treatment method for completely swelling into paste with cold water
Weigh an appropriate amount of modified starch and place it in a tall beaker.
The sample is moistened with anhydrous ethanol, and then slowly added with distilled or deionized water while stirring on the magnetic stirrer, the total volume is 2/3 of the beaker volume.
Stir for 10 minutes to ensure that the sample is evenly pasted without clumping.
Sample was taken for viscosity determination.
2. Prepare instruments
2.1 Adjustment and calibration
Make sure the measuring barrel of the rotating viscometer is level.
Calibrate and zero according to instrument instructions to ensure instrument accuracy.
2.2 Setting of thermal insulation device
The heat preservation device of the rotary viscometer is connected with the circulation device of the constant temperature water bath.
According to the needs of the experiment, the temperature of the constant temperature water bath is set to make the temperature of the measuring cylinder consistent with that of the water bath.
Step 3 Determine
The prepared sample is removed from a four-port flask or high beaker and added to the measuring barrel of the rotary viscometer.
Select the appropriate rotor according to the sample characteristics.
Set the appropriate speed for measurement.
Read and record viscosity values.
Matters needing attention
Ensure that the torque of the rotary viscometer is within its normal operating range. If out of range, need to adjust the speed and rotor.
According to the experimental requirements, the viscosity of the sample after different holding time can be determined.