LS-209-2 Two Box Thermal Shock Chamber

Kategoriler:

  • Açıklama
  • soruşturma

LS-209-2 Two Box Thermal Shock Chamber

Product nameTwo van hot and cold impact testing machine
Use of low temperature and high temperature hot and cold storage tank, in accordance with the cylinder, the hanging basket in turn to high temperature and low temperature box, so as to achieve rapid temperature shock effect, balance (BTC) + special temperature control system design of supply air circulation system, to control SSR P.I.D. way, make the system of heat equal to the amount of heat loss, thus the use of long-term stability.
Product modelTSD – 11 – a – 2 p (air-cooled type)
Not: air-cooled type need not install cooling towers, water cooled type to install, depends on customer’s requirement.
Limit sampleThe test equipment prohibited:

(1)Which is inflammable, explosive, easy to volatile substances sample test or store
(2) the corrosive substance sample test or store
(3) biological sample test or store
(4) strong electromagnetic emission source sample test or store

Ses, size and weight
Nominal content product52L
Shopping basket sizeW50*H32*D32cm
Installation dimensionsW132*H187*D155CM
ağırlıkAbout 400㎏
LS-209-2 Two Box Thermal Shock Chamber

Verim

Test environment conditionsEnvironment temperature is 25 ℃, relative humidity of 85% or less, test under the condition of no sample in the box
GB / T 5170.2 1996 temperature testing equipment.
High greenhouse+60°→+150°C
Time+60℃→+150℃ ≤25min
Sıcaklık aralığı-55℃~-10℃
Time+20℃ → -55℃≤60min
Ölçek (style)Test chamber upside for high temperature, lower for the low temperature zone, specimen is placed in the basket, basket turns in the high temperature box and low temperature in experiment
Shock temperature range(+60~+100)℃/(-40~-10)℃
Sıcaklık dalgalanma derecesi±0,5°C
The temperature deviation±2,0°C
The temperature recovery time≤5min
Temperature conversion time≤10sec
Restore conditionsHigh temperature exposure: + 150 ℃ / dakika – > low temperature exposure: 30 -e 55 ℃ / 30 dakika
Sample maximum weight: 2.5 KİLOGRAM
Meet the test standardGB / T 2423.1 2008 Ölçek: A test method at low temperature
GB / T 2423.2 2008 test B: high temperature experiment method
GB/T2423.22-2002 test N: temperature change test method test Na
GJB 150.5 Bir – 2009 military equipment laboratory test method for part 5: the temperature impact test
GJB 360 a – 1996 temperature impact test
GJB367.2-2001 military communication equipment general specification temperature impact test
IEC 68-2-14 test method for N: temperature changes
Structural features

Bize Ulaşın