What is a thermogravimetric analyzer? For many customers who have never touched this instrument, it may be relatively unfamiliar. However, for some industries engaged in the production of plastics, metals and plastics, the thermogravimetric analyzer is a commonly used testing instrument. It is an instrument that uses thermogravimetric method to measure the relationship between the mass of a substance and the temperature under programmed temperature control. The data measured by experiments are used to study the thermal stability and composition of different materials. In order to ensure the accuracy of the measurement data of the thermogravimetric analyzer, regular operation is required. The following is a small edition of Nanjing Dazhan Instrument to specifically understand the use method and operation precautions of the thermogravimetric analyzer.

What is the operation method of the thermogravimetric analyzer?

1, boot preheat. Especially at the beginning of the experiment, it needs to be turned on and preheated for about 3 hours.

2. Parameter setting. Through the touch screen of the instrument, adjust the temperature, heating rate and other parameters required by the experiment.

3. Blank experiment. Take two blank crucibles and place them on the test rod (when the experimental temperature is higher than 600℃, all ceramic crucibles must be used.) Cover the furnace, wait for 3 minutes, click the “Clear zero” on the instrument control screen, ensure that the values of “DSC” and “TG” are zero, and then proceed to the next operation.

4. Start experimenting. One of the crucibles is removed, and then the experimental sample is loaded into the instrument. Wait for 3 minutes, until the quality display on the instrument is very small, “Run” button.

5. Open the software. Open the software on the computer linked with the instrument, click “Start” to enter the test phase, and the experimental graph will appear on the software.

 

Thermogravimetric analyzer operation method and precautions

What are the precautions in the operation of the thermogravimetric analyzer?

1. Use a suitable crucible at high temperature. Thermogravimetric analyzer in the use of the process will have some supporting tools, such as: crucible. When doing the temperature test, if the test temperature reaches more than 600℃, it needs to be replaced with a ceramic crucible, because the ceramic crucible is more resistant to high temperature, which can ensure that the test sample will not be affected by the container, so as to ensure the accuracy of the test data.

2, strong acid and alkali samples need to be diluted in advance. If the thermogravimetric analyzer is used to test the samples of strong acids and alkali substances, it is necessary to pay attention to the dilution of the samples in accordance with the specified proportion before the next test can be carried out.

3, the sample quantity should meet the requirements. If the sample to be measured by the thermogravimetric analyzer is a liquid, it should be noted that the liquid level cannot exceed half of the crucible, and if it is a solid powder sample, it cannot exceed one-third. It is also important to ensure that the sample does not react with the crucible before testing.

Thermogravimetric analyzer operation method and precautions

4, high temperature test blank experiment. When the test temperature is high, the blank test should first be done on the crucible used with the thermogravimetric analyzer, the blank test curve should be adjusted as the baseline, and then the sample is put into the test. During the test, attention should also be paid to the indoor temperature on the operating panel of the thermogravimetric analyzer to avoid large fluctuations. At the same time, there should be no significant airflow, noise or vibration in the test environment.

5, regularly calibrate the thermogravimetric analyzer instrument. If the instrument is not used for a long time, it should be calibrated before the experiment. At the same time, the crucible used is a ceramic crucible, and it cannot be calibrated with a ceramic crucible during calibration.

6. After the experiment, turn off the power of the instrument. After the thermogravimetric analyzer experiment, the temperature of the instrument is finished, and the test sample is taken out, the test pattern is maintained, the power is turned off, and the instrument is placed in the original position.