Description

Thermal Storage Coefficient Measuring Instrument

Product Introduction

This thermal storage coefficient measuring instrument is primarily used to measure the thermal conductivity, thermal storage coefficient, specific heat, and thermal diffusivity of lightweight aggregate concrete, homogeneous slabs, lightweight aggregate concrete, colloidal, powdered, and granular materials, both dry and at varying moisture contents. It can simultaneously measure and calculate various thermophysical coefficients under both dry and equilibrium moisture content conditions in a single test.

 

The thermal storage coefficient measuring instrument offers simple operation, high accuracy, and fast test speed (one test takes approximately ten minutes). Based on the principle of unsteady heat conduction, it briefly heats the test material, causing the material’s temperature to change. Based on the characteristics of this temperature change, the thermal storage coefficient, thermal conductivity, thermal diffusivity, and specific heat can be calculated by solving the heat conduction differential equation.

 

This instrument is widely used by manufacturers of heat-resistant and thermal insulation materials, relevant quality inspection departments and units, universities, research institutes, and other scientific research institutions.

 

Standards

JGJ/T 12-2019 “Technical Specification for Lightweight Aggregate Concrete”

JG158-2004 “Glue Powder Polystyrene Granule External Wall Insulation System”

JG/T283-2010 “Expanded Vitrified Microsphere Lightweight Mortar”

Thermal Storage Coefficient Measuring Instrument

Technical Parameters

1. Thermal Conductivity Measurement Display Range  (0.001-10) W/(m•K)
2. Thermal Storage Coefficient Measurement Display Range  (0.01-100) W/(m2•K)
3. Relative Measurement Error  <5%
4. Temperature Measurement Range  (0-200)°C
5. Temperature Resolution  0.01°C
6. Thin Specimen Thickness  Range:(20-30) mm
7. Thick Specimen Thickness  Range:(60-100) mm
8. Specimen Dimensions  200 mm × 200 mm
9. Specimen Flatness  0.1 mm
10. Test Chamber Temperature  Standard Temperature (23 ± 2)°C
11. Test Chamber Humidity  Standard Humidity (40-60)% RH
12. Power Supply  AC 220V ± 10%, 200W