Description

Oxidation Stability Tester

The A1101 Oxidation Stability Tester (Metal Bath) is suitable for evaluating the oxidation stability of turbine oils with the same composition (base oil and additives). It can also be used to evaluate the oxidation stability of new mineral insulating oils containing 2,6-di-tert-butyl-p-cresol as a rapid method. The test results are used to verify the performance of new mineral insulating oils containing 2,6-di-tert-butyl-p-cresol, 2,6-di-tert-butylphenol, or both, allowing for continuous performance control for each batch of oil. It is mainly used in the petroleum, chemical, power, pharmaceutical, commodity inspection, scientific research, and environmental protection fields.

 

Standards

Applicable Standards: ASTM D 2272, ASTM D4742, SH/T0193

 

Feature

1. Suitable for new oil, in-use oil, steam turbine oil, lubricating oil, insulating oil, and automotive gasoline engine oil, etc. With appropriate accessories, it can also perform D2112 and D4742 tests;

2. The metal measuring tank reduces the unpleasant odor emitted by the medium oil during the test, avoiding potential dangers due to long test times;

3. The instrument can automatically perform detection, calculation, and control, requiring no personnel on-site during the entire testing process, with a high degree of automation. It automatically judges the experimental results;

4. Compact benchtop design, metal bath eliminates the need for liquid media. A manual valve protection device is added;

5. Simple ventilation, no need to place it in a fume hood, exhaust gases and odors can be easily discharged through plastic pipes;

6. Uses an Omron temperature controller, ensuring precise and stable temperature control and reliable experimental data;

7. Industrial touch screen all-in-one machine, using the WINDOWS operating system, with excellent data management capabilities.

 8. Can trace historical data curves from any period and store a large amount of historical data;

9. Can connect to a universal USB printer for A4-sized printing and printing historical curves;

10. Can perform simultaneous testing on two channels.

Oxidation Stability Tester

Technical Parameters

• Power supply  AC220V±10%
• Rated power  2500W
• Pressure sensor range  0~1.6MPa, accuracy: ±2‰
• Temperature control range  continuously adjustable from room temperature to 180.0℃, commonly used 150.0℃
• Temperature control accuracy  ±0.1℃
• Rotation mechanism speed  100±5r/min
• Angle between oxygen bomb and horizontal plane  30°
• Dimensions  490mm×450mm×640mm
• Instrument weight  30kg