MLC Series Electronic Analytical/Precision Balance Tester

Features

 supercell electromagnetic force sensor, is from the whole piece of high purity aluminum in high precision machining center integration on cutting machine parts fell 70%, the response speed increased by 30%, reduce the failure rate, so as to make the weighing results more accurate and reliable.

 to optimize the design of the product function

 

The machine adopts streamlined design, shortcut icon key navigation, beautiful appearance, convenient operation. With anti-leakage, anti-static, anti-corrosion properties, suitable for a variety of occasions.

 

The stainless steel scale plate makes the weighing articles stable and ensures the accuracy of weighing.

 

 5 inch colour touch screen

 

5 inch color touch screen, easy to read weighing information, intuitive operation interface and quick icon menu navigation, sensitive and convenient operation.

 

 transparent glass wind shield

 

The sample can be seen in all directions during the weighing process, and can also be disassembled and cleaned to realize fast cleaning.

 

 a variety of weighing mode

 

Basic weighing, percentage weighing and counting weighing modes.

 

 various weighing unit conversion functions

 

Units such as grams (g), carats (Ct), ounces (oz) and pounds (lb) can be switched freely.

 

 built-in RS232 bi-directional communication interface

 

Can realize the balance and computer, micro printer connection. The weighing results can also be transmitted to other open applications such as PC.

 

 automatic calibration system, to ensure accurate weighing

 

When the temperature changes by 1.5℃ or more than 2 hours since the last calibration time, the balance will automatically start the calibration function, avoiding the potential factors of untimed calibration or inaccurate external weights, making the user’s weighing results more accurate and reliable, and reducing unnecessary operations (internal calibration balance only).

 
MLC Series Electronic Analytical/Precision Balance Tester
Parameters
ModelMLC124MLC224MLC124AMLC224A
Range120g220g120g220g
readability0.1mg
repeatability(≤)±0.1mg
linear error(≤)±0.2mg
CalibrationExternal weight calibrationFully automatic internal calibration
operating temperature range10℃~30℃
Operating humidity range10%RH~70%RH
Response time (mean)2.5 second
Sample plate sizeΦ90mm
Dimensions210mm×310mm×325mm
Height of weighing room210mm
Startup preheating time30-60 second
Dc adapterinput:220V AC/50HZ;output:12V DC/2A
Baud rate300、600、1200、2400、4800、9600
expertise ratingⅠ  lever