LS-8270 XRD X-ray Diffractometer
- Description
- Inquiry
Description
LS-8270 XRD X-ray Diffractometer
Overview
LS-8270 mini benchtop X-ray diffractometer is a multipurpose powder diffraction analytical
instrument that specifically designed and developed for industrial production and quality control. LS-8270mini offers a lot – in a highly compact design, especially suitable for catalyst, titanium dioxide, cement, pharmaceutical and other product manufacturing industries.
Application
1.Phase identification in unknown samples ·
2.Quantitative analysis of known phases inmixed samples
- Rievtveld analysis
Attachments
1.Graphite monochromator
2.DCX-6 Sample changer
3.In situ cell attachment
LS-8270 XRD X-ray Diffractometer
Technical parameter
| Power | Solid state X-ray generator: 600W(40kV, 15mA).
Reliability: <0.005% |
|
| X-ray tube | Cermet X-ray tube, Cu target, power: 2.4kW, focus: 1 x 10mm
Water cooling (water flowrate>2L/min) |
|
| Goniometer | Sample horizontal θs-θd structure | |
| Diffraction circle radius | 150mm | |
| Measurement range | -3 ~150° | |
| Scan mode | Stepping, Omg | |
| Minimum step width angle | 0.0001° | |
| Angle repeatability | 0.0005° | |
| Linearity of diffraction angle | <0.02° (Standard sample, within the full spectrum) | |
| Detector | Closed proportional detector | Semiconductor array detector |
| Maximum linear count rate | 5×105CPS | 9×107CPS |
| Energy resolution | =25% | =1keV |
| Scatter dose | =1μSv/h (Outside the X-ray protective device) | |
| Comprehensive stability | =0.5% | |
| Dimensions | 640×480×770 (W×D×H) mm | |
Features
- High-frequency high-voltage X-ray generator does not only improve the stability of the instrument, but
also ensure the repeatability of the measured data.
- The goniometer θs and θd arms are driven by servo motor + optical coding control technology. The diffrac
tion angle linearity is less than 0.02° in the diffraction angle measurement range.
- Scattered ray protection is safer and more reliable; the radiation protection door is automatically prohibi
ted from opening when the sample is measured. In any case, operators can be protected from scattered
radiation.
- Configure the most advanced metal ceramic X ray tube, to ensure long service life.
- Compact size can be mounted on the bench without the need for a specific laboratory environment. Simple to use, operate and maintain.