Concrete thermophysical parameter tester
- Description
- Inquiry
Description
Concrete thermophysical parameter tester
Description:
The Concrete Thermophysical Parameter Tester is suitable for measuring concrete thermal parameters such as adiabatic temperature rise, specific heat capacity, thermal diffusivity (thermal diffusivity, thermal conductivity), and thermal linear expansion coefficient. The instrument also features ambient temperature simulation and a constant temperature test chamber.
Product Introduction
The Concrete Thermophysical Parameter Tester is used to measure the adiabatic temperature rise of concrete, specific heat, thermal conductivity, thermal conductivity, and linear expansion coefficient of concrete (five tests are available, with some optional test items available).
Standards
DL/T5150-2001 “Test Procedure for Hydraulic Concrete”
JG/T 329-2011 “Concrete Thermophysical Parameter Meter”
GB/T50080-2016 “Test Method for New Energy of Ordinary Concrete Mixtures”
GB/T50081-2019 “Standard for Test Methods for Thermophysical and Mechanical Properties of Concrete”
JTG3420-2020 “Test Procedure for Cement and Cement Concrete for Highway Engineering”
T0539-2020 “Test Method for Adiabatic Temperature Rise of Cement Concrete Mixtures”
Concrete thermophysical parameter tester
Technical Parameters
1. Adiabatic Temperature Rise Measurement
Test Temperature: 5°C to 80°C
Adiabatic Temperature Tracking: Less than ±0.05°C (10°C to 80°C)
Temperature Variation (72-Hour Tracking with 50L of Water Without a Heat Source): Less than ±0.05°C
Adiabatic Temperature Rise Characteristics: In typical adiabatic temperature rise design methods, in addition to meeting national standards for tracking accuracy of ≤±0.05°C, in order to ensure consistency between the ambient temperature and the specimen temperature, a deviation value corresponding to the specimen temperature must be calibrated to ensure the authenticity of the concrete specimen’s adiabatic temperature rise. Even with these measures, the randomness of system errors often prevents a one-to-one correspondence between the deviation value and the tracking accuracy. Consequently, the concrete specimen may release or absorb more or less heat into or out of the environment, thus rendering the concrete specimen temperature curve less accurate. To address this issue, our company has designed a software and hardware control system based on these two measures. This reduces the amount of heat released or absorbed by the specimen into or out of the environment by half, while maintaining the same tracking accuracy and deviation value.
2. Specific Heat Capacity Measurement
Test Temperature: 10°C to 75°C
Adiabatic Temperature Tracking: Less than ±0.2°C
3. Thermal Diffusivity (Thermal Diffusion) Measurement
|
Test Temperature: 5°C to 75°C Low Thermal Diffusivity Temperature: Initial Specimen Temperature +15°C Constant Temperature Variation: Less than ±0.3°C, ±0.05°C during Stable Thermal Diffusivity |
4. Thermal Expansion (Linear Expansion) Coefficient
Test Temperature: 5°C to 75°C
Constant Temperature Variation: Less than ±0.1°C
5. High-Precision Constant Temperature Test Chamber
Test Temperature: 5°C to 80°C
Constant Temperature Accuracy: Less than ±0.1°C (10°C to 80°C)