Bauer Fiber Sieve Analyzer
Product Overview:
The Bauer Fiber Sieve Analyzer is an instrument used in pulp and paper laboratories to determine the weight-average fiber length of pulp fibers using the sieving method. It measures the proportion of fibers of different lengths in pulp, revealing the distribution of fibers of various lengths in the pulp. This machine conforms to the sieving test method specified in GB/T 2678.1-1993, as well as the sieving test method specified in TAPPI T233cm-95, and meets the Nordic SCAN M6 test method (10g sample, water flow for 15 min, flow rate 10L/min).
Principle:
The pulp suspension is injected into the sieving device. Each sieving container is equipped with a sieve, arranged in a stepped manner. As the pulp suspension flows from one container to another, pulp of different fiber lengths is retained on different sieves. The fiber length of the retained pulp matches the mesh size of the sieve plate. After sieving for a certain period of time at a certain water flow rate… The fibers are collected from each sieve plate, dried, and weighed to a constant weight. The result is reported as a percentage of the mass of the sample based on the amount of fiber remaining on each mesh.
Standard Reference:TAPPI T233cm-95, GB/T 2678.1,SCAN M6
Main Applications:
The Pall fiber sieve analyzer is mainly used to measure the proportion of fibers of different lengths in pulp, helping users understand the distribution of fibers of various lengths in the pulp. This is an important tool for the pulp and paper industry to understand pulp quality and optimize production processes.
Working Principle:
The working principle of the Pall fiber sieve analyzer is to determine the length of fibers by grading. The instrument consists of multiple stepped containers, each equipped with a stirrer and replaceable sieves. The sieve size decreases sequentially to ensure that fibers of different lengths can be effectively separated.
During the experiment, pulp suspension is injected into the sieve separator. As the pulp suspension flows from one container to another, fibers of different lengths are trapped on different sieves. After a certain sieving time, the fibers on each sieve are collected, dried, and subjected to constant weight treatment. The results are then reported as a percentage of the sample mass based on the amount of fiber retained on each mesh.
Functional Features:
Multi-stage sieving: The instrument features multiple stepped containers and replaceable sieves, enabling multi-stage sieving for more accurate measurement of the distribution of fibers of different lengths in the pulp.
Automatic control: The Pall fiber sieve separator employs fully automatic control, is simple to operate, and easy to learn. The integrated timer allows users to set individual grading cycles, ensuring the accuracy of experimental results.
Standard Compliance: This instrument conforms to the sieving test method specified in GB/T 2678.1-1993, as well as the sieving test method specified in TAPPI T233cm-95, and meets the Nordic SCAN M6 test method (10g sample, water flow for 15 min, flow rate 10L/min).
Easy to Clean: The instrument is designed with a water injection device and a suction pump, allowing for quick and easy cleaning after sieving, facilitating future use.
Superior Materials: The instrument is primarily made of stainless steel, possessing excellent corrosion resistance and durability, ensuring long-term stable operation.
Application Scenarios:
The Bauer fiber sieve analyzer is widely used in pulp and paper laboratories, paper quality testing centers, and related research institutions. It helps users quickly and accurately understand the distribution of fiber length in pulp, providing strong support for optimizing production processes and improving product quality.
Bauer Fiber Sieve Analyzer
Technical Parameters
|
Filter tray material: |
304 Stainless Steel |
External Dimensions: |
1780mm (L) × 520mm (W) × 1680mm (H) |
|
Net weight: |
130Kg |
Water Source Conditions: |
Requires continuous water supply or independent white water circulation. |
|
Power supply and capacity: |
220V/1000W |
Commonly Used Screens (Recommended): |
Long fibers, Tyler sieve sizes: 10 mesh, 14 mesh, 28 mesh, 48 mesh, 200 mesh. |
|
Common screen (recommended): |
Medium-length Fiber, Tyler sieve sizes (factory selected): 14 mesh, 28 mesh, 48 mesh, 100 mesh, 200 mesh |
Commonly Used Screens (Recommended): |
Short fibers, Tyler sieve sizes: 28 mesh, 48 mesh, 100 mesh, 150 mesh, or 200 mesh. |
|
Specialty fiber application: |
10 mesh, 14 mesh, 28 mesh, 30 mesh, 40 mesh, 60 mesh, 80 mesh, 100 mesh, 200 mesh, 300 mesh, 400 mesh, 500 mesh, 1000 mesh, 3000 mesh. The mesh size must be selected experimentally based on the specific functional fiber characteristics. |
|
|
|
Screening grades: |
Grade 4 (standard), Grade 5, and Grade 6 are available upon request. |
Screening mesh: |
(Common plant fibers): Tyler sieve sizes 14 mesh, 28 mesh, 48 mesh, 100 mesh; Special fibers require special configuration based on fiber characteristics. |
|
Screening device capacity: |
10 liters (capacity of a single classifier) |
Agitator speed: |
(580±40) r/min |
|
Water flow rate: |
2~18 L/min adjustable (adjustable according to experimental standards) |
White water circulation: |
Customized configuration, mainly used for special fibers with high screening difficulty and long screening time, primarily high-molecular synthetic fibers. |
|
Screening time: |
0~9999 min adjustable (adjustable according to experimental standards) |
Filtration efficiency: |
R0.50mm/97H/cm2 |
|
Frame material: |
304 stainless steel |
Vacuum source: |
5-channel water ring is vacuum, multi-stage independent completion (customized for special fibers) |
|
Slurry recovery: |
Vacuum and non-vacuum switching; quick installation and rapid opening of the material chamber (customizable for special fibers) |
Pulp tank material: |
304 stainless steel |
Package Contents:
Bauer Fiber Sieve (5 stages) ×1
Dedicated Sieve Plate ×1
Water Spray Gun and Water Hose ×1
High-Pressure Water Hose ×1
Instruction Manual ×1
Warranty Card and Certificate of Conformity ×1