Benutzen

The hot air aging test box is suitable for the heat resistance test of electrical insulation materials, the air exchange aging test of electronic spare parts, plasticized products, etc.The adaptability of the sample to storage and use in high temperature environment was evaluated and judged. The performance of the sample after aging in air under simulated high temperature and atmospheric pressure was measured and compared with the performance of the non-aging sample.

Teststandard:In accordance with the standard GB/ t3512-2014 /ISO 188 2011 “vulcanized rubber or thermoplastic rubber/hot air accelerated aging and heat resistance testturbulence design.

 

LS-214 Ventilation Aging Chamber

Enclosure structure

  1. Interne Größe: 600(Eine neue Batterie kann länger als ununterbrochen verwendet werden)×600(depth)×650(Eine neue Batterie kann länger als ununterbrochen verwendet werden)mm;
  2. External dimension: 1300(Eine neue Batterie kann länger als ununterbrochen verwendet werden)×850(depth)×1700(Eine neue Batterie kann länger als ununterbrochen verwendet werden)mm, subject to the actual object;
  3. Operation panel: the control panel is located on the right side of the machine (as shown in the figure);
  4. Opening mode: single door is opened from right to left;
  5. Observation window: toughened glass window on the door, size W250*H350mm;
  6. Inner material: 304# stainless steel sand board: thick 1.2mm;
  7. External material: cold-rolled steel plate, 1.0mm thick powder baking paint treatment;(beige)
  8. Product materials: two built-in turntables with a diameter of 300mm;
  9. One set of speed-regulating motors with adjustable speed of 3-20 Drehzahl;
  10. Thermal insulation material: high temperature resistant fiber rock wool, good thermal insulation effect;

11, the door sealing material is: high temperature resistant silicone strip;

  1. The machine is equipped with movable casters and fixed feet, which can be pushed and fixed at any time.

 

LS-214 Ventilation Aging Chamber

Temperature control system

  1. Temperature control is PID digital display instrument (Taiwan taisong instrument GB705), single point temperature control, automatic calculation, simultaneous display of PV/SV, button setting;
  2. Timing function: from 1 zu 9999 (hour, minute and second can be set), the temperature reaches the timer, when the time arrives, the heating and power supply will stop, and the alarm will be given;
  3. Temperature line (K) Art;
  4. The output is 3-32v;
  5. The current controller is: SSR solid state relay;
  6. Heating material: stainless steel high-temperature heating tube;
  7. Temperaturbereich: adjustable from normal temperature +15 to 300℃;
  8. Heating rate: Über 18 minutes at 100℃;
  9. Control accuracy: ± 0,5 ℃;
  10. Display accuracy: 0.1℃;
  11. Temperature deviation: RT+15 ~ 200℃±2.0℃, 201 ~ 300℃±3.0℃ (no-load test)

 

LS-214 Ventilation Aging Chamber

Ventilation System

  1. Air ventilation: 3-10 times/hour;
  2. Power meter for testing when air is changed or not changed;
  3. Air ventilation is controlled by adjustable air duct, and the size of new air is manually adjusted for natural air ventilation;
  4. Manual adjustable air duct is used for air discharge, which is located at the top of the machine;(exhaust should be discharged to the outside)
  5. Ventilation: natural air intake;
  6. The air outlet is located at the top of the machine, and the manual switch tuyere is adopted;

 

LS-214 Ventilation Aging Chamber

Circulating air supply system

  1. Air supply: internal hot air circulation;
  2. The motor is special for long shaft with high temperature resistance, with a power of 370W/220V;
  3. Wind wheel multi-wing turbine type 9 “;

Safety protection system

  1. Over-temperature protection system: when the temperature is out of control, the heating and power supply will stop automatically when the over-temperature protection temperature is exceeded, so as to protect the safety of the product and the machine;
  2. Circuit protection system: ground protection, quick insurance, usw.

 

LS-214 Ventilation Aging Chamber

Power supply unit

  1. The voltage is 220V
  2. Power 4.5kw
  3. Current 21A

Kontaktiere uns