Açıklama

TGA DSC fonksiyonlu STA Eşzamanlı Termal Analiz Cihazı

The operating principle of a simultaneous thermal analyzer (STA) is to measure the change in mass of a substance as a function of temperature or time, using programmed temperature control. Its main components include a heating furnace, a balance, a temperature controller, and a recording system.

Operating Principle

A simultaneous thermal analyzer operates based on the temperature changes of a sample and the processes of heat conduction, emme, and radiation. It controls the temperature of the furnace to heat, cool, and maintain a constant temperature for the sample, then monitors and records the sample’s temperature changes and the corresponding physicochemical properties.

Başvuru

A simultaneous thermal analyzer can simultaneously measure both TG and DSC information, covering a wide range of measurement areas, including thermal stability, redox analysis, decomposition behavior, corrosion studies, decomposition kinetics, melting/crystallization, solid-state phase transitions, kristallik, and glass transition.

Simultaneous thermal analyzers are widely used in research and development, process optimization, and quality control in various fields, içermek plastikler, silgi, kaplamalar, eczacılık, katalizörler, inorganik malzemeler, metallic materials, ve kompozit malzemeler.

Application area

Widely used in research and development, process optimization, and quality monitoring in various fields such as plastics, silgi, kaplamalar, eczacılık, katalizörler, inorganik malzemeler, metal materials, ve kompozit malzemeler.

 

Structural Advantages

1. The furnace heating system utilizes double-wound precious metal platinum-rhodium alloy wire to reduce interference and enhance high-temperature resistance.

2. The tray sensor is precision-crafted from precious metal alloy wire, offering high-temperature, oksidasyon, ve korozyon direnci.

3. The power supply and heat dissipation circuits are separated from the main unit to minimize the impact of heat and vibration on the microthermobalance.

4. The main unit utilizes an insulated heating furnace to minimize the thermal impact on the chassis and microthermobalance.

5. The furnace utilizes double insulation for improved linearity; the furnace features automatic lift for rapid cooling; and an exhaust outlet allows for integration with infrared sensors.

Controller and Software Advantages:

1. Utilizes an imported ARM processor for faster sampling and processing speeds.

2. Four-channel sampling AD acquires TG and temperature signals.

3. Heating control utilizes a PID algorithm for precise control. Multi-stage heating and constant temperature control are possible.

4. USB bidirectional communication between the software and the instrument enables complete remote operation. Instrument parameters can be set and stopped via the computer software. 5. 7-inch full-color 24-bit touchscreen for an enhanced human-machine interface. TG calibration can be performed directly on the touchscreen.

Compliant with industry standards:

ASTM D 3386, ASTM D 696, ASTM E831-2019, ASTM E831-06, ASTM E831-2014, DIN 51045-3-2009, GB / T 33047.3-2021, GB / T 4498.2-2017, GB / T 32868-2016, GB / T 31984-2015, GB / T 14837.2-2014, GB / T 29189-2012, vesaire.

TGA DSC fonksiyonlu STA Eşzamanlı Termal Analiz Cihazı

Teknik parametre

Sıcaklık aralığı

Room temperature -1500°C(Equipped platinum heating wire)

Sıcaklık çözünürlüğü

0.001°C

Sıcaklık dalgalanması

±0,01°C

Heating/cooling rate

0.1~100°C/min; -0.1~-40°C/min

Sıcaklık kontrol yöntemi

PID kontrolü, ısıtma, soğutma, and constant temperature

Weighing system measurement range

0.001mg~3g, expandable to 50g

Doğruluk

1μg

Duyarlılık

0.1μg

DSC range

±1000mW

DSC sensitivity

0.1µW

Atmosfer kontrolü

Built-in two-way gas flowmeter, including switching between two gas lines and flow rate control

Yazılım

Intelligent software automatically records TG curves for data processing, with TG/DTG, yığın, and percentage coordinates freely switchable; software features automatic scaling and extension based on the graph display

Hot-hold time

Any settings are available; standard configuration ≤600 minutes

The operating software comes with a copyright certificate, and data testing frequencies can be selected in real time, 2s, 5s, or 10s.

The furnace features both automatic and manual lift modes, enabling rapid temperature reduction; from 1000°C to 50°C in ≤ 20 dakika.

An external water cooling device isolates heat from drifting the weighing system; the temperature range is -10°C to 60°C.