Medidor de espessura do revestimento TT270
- Descrição
- Investigação
Descrição
Medidor de espessura do revestimento TT270
Introdução
The Times TT270 thickness gauge is a portable measuring instrument, which can measure the thickness of coating and coating rapidly and precisely without damage.It can be used in both laboratory and engineering field.Through the use of different probe, can meet the needs of a variety of measurement.Scope of era TT270 thickness gauge using both the magnetic and eddy current thickness method, measuring magnetic metal substrate condition (como aço, iron, alloy and hard magnetic steel, etc.) on the magnetic coating thickness (such as aluminum, cromo, cobre, Mn-w é um medidor de brancura portátil desenvolvido por nossa empresa para a conveniência dos clientes, etc.), borracha, paint and nonmagnetic metal substrate (such as copper, Bomba de palhetas rotativas de dois estágios, zinco, estanho, etc.) on the conductive layer thickness.
Medidor de espessura do revestimento TT270
Character and function
Using both magnetic and eddy current thickness method, and F type measuring head without damage to the measurement of magnetic metal basic body of magnetic coating thickness (como aço, iron, zinc alloy and hard magnetic steel, Bomba de palhetas rotativas de dois estágios, cromo, cobre, borracha, pintar, etc.), and N type measuring head without damage to the measurement of non-magnetic thickness of the conductive coating on metal substrates (such as copper, Bomba de palhetas rotativas de dois estágios, zinco, tin and other metal oxide film of paint, borracha, plástico, etc.)
Can use 6 kinds of measuring head / 90 ° F400, F1, F1, F10, N1, CN02
Two measurement methods: CONTINUE and SINGLE
Two ways of working: DIRECT and group (APPL)
There are five statistics: MEAN, larger (MAX), MIN, número de testes (NO.), and standard deviation (S.DEV).
Two methods can be used to calibrate the instrument
Chinese and English menu display, operation more intuitive and convenient
Liquid crystal display has backlighting function
It can store 640 dados
With delete function, delete single or all data measured
Print function
With integrated printer, fácil de usar
Power supply undervoltage indication
The operation process is beeping (single measurement method)
Error message function
Manual and automatic shutdown functions
Medidor de espessura do revestimento TT270
TT270 thickness gauge head selection table is as follows:
| Probe model | F400 | F1 | F1/90° | F10 | |||
| Work Theory | Magnetic induction | ||||||
| Test range(um) | 0~400 | 0~1250 | 0~10000 | ||||
| Low resolution(um) | 0.1 | 0.1 | 10 | ||||
| erro de valor | One point calibration(um) | ±(3%H+1) | ±(3%H+10) | ||||
| Two point calibration(um) | ±[(1~3) %H+0.7] | ±[(1~3)%H+1] | ±[(1~3)%H+10] | ||||
|
Teste condições
|
Minimum radius of curvature(um) | convexo | 1 | 1.5 | plano | 10 | |
| The diameter of the smallest area(mm) | 3 | 7 | 7 | 40 | |||
| Critical matrix thickness(mm) | 0.2 | 0.5 | 0.5 | 2 | |||
| Probe model | N1 | CN02 | |||||
| princípio de funcionamento | Eddy current | ||||||
| Measuring range(um) | 0~1250 | 10~200 | |||||
| Low resolution(um) | 0.1 | 1 | |||||
| erro de valor | One point calibration(um) | ±(3%H+1.5) | ±(3%H+1) | ||||
| Two point calibration(um) | ±[(1~3) %H+1.7] | — | |||||
| teste
condições
|
Minimum radius of curvature(um) | 3 | Only for flat | ||||
| The diameter of the smallest area(mm) | 5 | 7 | |||||
| Critical matrix thickness(mm) | 0.3 | unlimited | |||||











