Descrição

Analisador de óxido de cálcio livre totalmente automático

Introdução do produto

This fully automatic free calcium oxide analyzer was developed and manufactured by our company in accordance with GB/T 176-2017, “Cement Chemical Analysis Methods – Determination of Free Calcium Oxide – Ethylene Glycol Method.” This device is primarily used to measure the free calcium oxide content in clinker and cement.

Princípio de funcionamento

The free calcium oxide in the sample reacts with an ethylene glycol-anhydrous ethanol solution under heating to form a weakly alkaline calcium glycolate, which causes the phenolphthalein indicator to turn purple-red. The TSWL-YL100 fully automatic free calcium oxide analyzer automatically controls the heating temperature and time to convert all free calcium oxide into calcium glycolate. During the subsequent titration process, the instrument uses a color sensor to observe the solution color and titrates the test solution with a standard benzoic acid anhydrous ethanol solution until the red color disappears. The free calcium oxide content in the sample can be calculated based on the volume of benzoic acid anhydrous ethanol solution consumed until the red color disappears, the sample mass, and the titrant.

Analisador de óxido de cálcio livre totalmente automático

Parâmetros técnicos

1. Faixa de medição: 0-10% (f-CaO Content)

2. Tempo de análise: 10 min

3. Tamanho da amostra: 0.5 g

4. Repeatability Limit: 0.14% when f-CaO ≤ 2%

0.19% when f-CaO > 2%

5. Incluindo vários tipos de tubos cosméticos: 30-100°C

6. Number of Samples: Dois

7. Método de controle: Tela sensível ao toque

8. Poder: 1 kW

9. Incluindo vários tipos de tubos cosméticos: 220V 50Hz

Requisitos Ambientais

1. Temperatura Ambiente: 10°C-30 °C, Relative Humidity ≤ 85%

2. Nenhuma vibração forte, pó, forte interferência eletromagnética, ou gases corrosivos.

3. The laboratory room temperature should be kept as constant as possible.

4. The room should be free of strong air convection, fontes de calor fortes, fans, etc.

5. Avoid sunlight. The instrument should be placed in a place not exposed to direct sunlight.