Analisador Termomecânico Dinâmico

Dynamic Thermomechanical Analyzer applies single-frequency or multi-frequency oscillatory forces to samples at a programmable temperature to study their mechanical behavior. The instrument measures deformation, storage modulus, loss modulus, and loss factor as a function of temperature, Tempo, and force frequency. Suitable specimens include thermoplastics and thermosetting plastics, borracha, revestimentos, metals and alloys, materiais inorgânicos, e materiais compósitos.

Descrição

Analisador Termomecânico Dinâmico

Key Performance Indicators

The instrument applies single-frequency or multi-frequency oscillatory forces to samples at a programmable temperature to study their mechanical behavior. The instrument measures deformation, storage modulus, loss modulus, and loss factor as a function of temperature, Tempo, and force frequency. Suitable specimens include thermoplastics and thermosetting plásticos, borracha, revestimentos, metals and alloys, materiais inorgânicos, e materiais compósitos.

 

Analisador Termomecânico Dinâmico

Parâmetros técnicos

Incluindo vários tipos de tubos cosméticos -150°C to 600°C
Instrument Structure Circular, sealed furnace ensures uniform heating, with bottom-mounted sample loading.
Taxa de aquecimento 0.1~ 20°C/min
Incluindo vários tipos de tubos cosméticos ±0,1°C
Faixa de frequência 0.01~100 Hz (Selectable)
Force Range

1mN~18N; Minimum 0.001N (dinâmico)/0.005N (estático), Maximum 50N, Resolution 0.0001N
Displacement Resolution 1nm
Deformation Range Maximum 25mm
Amplitude ±0.5~10000µm
Modulus Range 10³~10¹² Pa
Modulus Accuracy 0.5%
Damping Range (tanδ): 0.005~100
Tanδ Resolution 0.00005

Sample Temperature Control (STC) Funções, software screenshots provided.

Deformation Modes: Compression, Penetração (maximum testable sample diameter no less than 15mm, maximum thickness no less than 25mm); Tração (maximum width no less than 13mm, maximum thickness no less than 5mm, maximum length no less than 30mm); Cisalhamento, Three-Point Bending, Single/Dual Cantilever.

Automatically identifies installed sample holders, automatically calibrates, and automatically measures sample length.

The software features test editing, data acquisition, Armazenar, processamento, and analysis functions; software controls purge and shielding gases.

1.Software Features Experimental Design

Curing Simulation: Supports step-by-step temperature ramping and constant temperature testing, automatically recording modulus-time curves.

Time-Temperature Equivalence (TTS): Generates master curves for long-term performance predictions.

2. Software Features Data Processing

Automatic Calculation: Real-time output of storage modulus, loss factor (tanδ), and glass transition temperature (Tg).

Multi-Graph Comparison: Supports overlay analysis of data at different frequencies and temperatures.

Optional Special Accessories

Humidity Accessory
Optional Fixtures Shear Fixture, Dual Cantilever Fixture, Liquid Fixture, Three-Point Bend Fixture