LS-215-1 Xenon lamp aging test chamber

LS-215-1 Xenon lamp aging test chamber

  • Description
  • Inquiry

LS-215-1  Xenon lamp aging test chamber

1.Profile.

Xenon lamp aging test chamber simulates the full spectrum lamp with Xenon arc lamp to reappear the destructive wave. It serves for the environmental simulation and accelerated test for research and development, product development and quality control. Besides, the chamber also applies to the choice of new material, weathering test after composition of material and improvement for the current material. It simulate different environment, and show the material changes under the explosion of sunlight.

The chamber meets the standard: GB/T16585-1996 , GB14522-93 , GB/T16422.3-97,ASTM G-155  ASTM D2565 and so on.

The Chamber is to determine the color fastness, aging resistance of textiles, plastics, coatings, rubber products or other materials by objecting to the simulated conditions of nature weather (sun, rain, temperature and humidity, etc.).

 

LS-215-1 Xenon lamp aging test chamber

1.Material of construction
u Full spectrum lamp

u Optional filter system

u Water spraying function

u Humidity control

u Temp. control system in the chamber

u Sample fixed hold of irregular shape

u USA imported xenon arc lamp

u Easy installation and operation, seldom daily maintainance.

u Service life time deepens on the irradiance level, usually 1600hrs. It’s easy to replace the lamp, with

filter of long service life time to ensure the necessary spectrum.

2.Technical parameters.
Model No., equipment nameLS-215-1-150 Xenon aging test chamber
Interior sizesW500×D500×H600mm
Exterior sizesW1050×D1050×H1750mm (or so)
Temperature range0℃~80℃
Controllerimported programmable 7inch touch screen
Machine components materialinner and outer 304# stainless steel
Sample holderAluminum alloy frame base
Protection deviceresidual-current circuit breaker (RCCB) to control the warning for loop overload, thermal protection and water shortage protection.
Temp. deflection±2℃  (without lighting)
Humidity range30%~98%R.H
Humid. deflection±2.5%R.H(Humid.≥75%R.H); ±3%R.H(Humid.≤75%R.H) without lighting
Light sourceimported full spectrum xenon arc lamp(average useful life 1600hrs)
Raining time1~9999min (adjustable)
Raining period1~240min, interval(cut)adjustable
lamp powerTest period for Temp., irradiation, condensation, spray can be adjustable
Lamp1.8KW
Wave length range290-800nm
heating power (or so)4.0KW
humidification (or so)2.0KW
Light irradiance≤1200w/m2
Distance between arc center and sample holder:350~380mm
Revolving speed of sample holder≥1r/min
water spray perioda. spray time: 0~99h59min (Continuing or adjustable)

b. stop time: 0~99h59min, (Continuing or adjustable)

lighting timeContinuing programming control available
black panel temp.85±3℃
cooling methodMechanical wind cooling; Room temp. less than 26℃ is recommend, relative humid: ≤85%R.H; humid. and temp. tested after 2 hours under idle condition.