LS-8270 XRD X-ray Diffractometer
- Opis
- Zapytanie ofertowe
Opis
LS-8270 XRD X-ray Diffractometer
Przegląd
LS-8270 mini benchtop X-ray diffractometer is a multipurpose powder diffraction analytical
instrument that specifically designed and developed for industrial production and quality control. LS-8270mini offers a lot – in a highly compact design, especially suitable for catalyst, titanium dioxide, cement, pharmaceutical and other product manufacturing industries.
Podanie
1.Phase identification in unknown samples ·
2.Quantitative analysis of known phases inmixed samples
- Rievtveld analysis
Załączniki
1.Graphite monochromator
2.DCX-6 Sample changer
3.In situ cell attachment
LS-8270 XRD X-ray Diffractometer
Parametr techniczny
| Moc | Solid state X-ray generator: 600W.(40kV, 15mama).
Niezawodność: <0.005% |
|
| X-ray tube | Cermet X-ray tube, Cu target, moc: 2.4kW, focus: 1 x 10mm
Chłodzenie wodne (water flowrate>2L/min) |
|
| Goniometer | Sample horizontal θs-θd structure | |
| Diffraction circle radius | 150mm | |
| Zakres pomiaru | -3 ~150° | |
| Scan mode | Stepping, Omg | |
| Minimum step width angle | 0.0001° | |
| Angle repeatability | 0.0005° | |
| Linearity of diffraction angle | <0.02° (Próbka standardowa, within the full spectrum) | |
| Detektor | Closed proportional detector | Semiconductor array detector |
| Maximum linear count rate | 5×105CPS | 9×107CPS |
| Energy resolution | =25% | =1keV |
| Scatter dose | =1μSv/h (Outside the X-ray protective device) | |
| Comprehensive stability | =0.5% | |
| Wymiary | 640×480×770 (Szer.×Głęb.×Wys) mm | |
Cechy
- High-frequency high-voltage X-ray generator does not only improve the stability of the instrument, but
also ensure the repeatability of the measured data.
- The goniometer θs and θd arms are driven by servo motor + optical coding control technology. The diffrac
tion angle linearity is less than 0.02° in the diffraction angle measurement range.
- Scattered ray protection is safer and more reliable; the radiation protection door is automatically prohibi
ted from opening when the sample is measured. W każdym razie, operators can be protected from scattered
promieniowanie.
- Configure the most advanced metal ceramic X ray tube, aby zapewnić długą żywotność.
- Kompaktowy rozmiar można zamontować na stole warsztatowym bez konieczności tworzenia specjalnego środowiska laboratoryjnego. Prosty w użyciu, obsługiwać i konserwować.