Meter Ketebalan Salutan TT270
pengenalan
The Times TT270 thickness gauge is a portable measuring instrument, which can measure the thickness of coating and coating rapidly and precisely without damage.It can be used in both laboratory and engineering field.Through the use of different probe, can meet the needs of a variety of measurement.Scope of era TT270 thickness gauge using both the magnetic and eddy current thickness method, measuring magnetic metal substrate condition (seperti keluli, besi, alloy and hard magnetic steel, dan lain-lain.) on the magnetic coating thickness (such as aluminum, chromium, tembaga, enamel, dan lain-lain.), getah, paint and nonmagnetic metal substrate (such as copper, aluminium, zink, timah, dan lain-lain.) on the conductive layer thickness.
Meter Ketebalan Salutan TT270
Character and function
Using both magnetic and eddy current thickness method, and F type measuring head without damage to the measurement of magnetic metal basic body of magnetic coating thickness (seperti keluli, besi, zinc alloy and hard magnetic steel, aluminium, chromium, tembaga, getah, cat, dan lain-lain.), and N type measuring head without damage to the measurement of non-magnetic thickness of the conductive coating on metal substrates (such as copper, aluminium, zink, tin and other metal oxide film of paint, getah, plastik, dan lain-lain.)
Can use 6 kinds of measuring head / 90 ° F400, F1, F1, F10, N1, CN02
Two measurement methods: CONTINUE and SINGLE
Two ways of working: DIRECT and group (APPL)
There are five statistics: MEAN, larger (MAX), MIN, number of tests (NO.), and standard deviation (S.DEV).
Two methods can be used to calibrate the instrument
Chinese and English menu display, operation more intuitive and convenient
Liquid crystal display has backlighting function
It can store 640 data
With delete function, delete single or all data measured
Print function
With integrated printer, mudah digunakan
Power supply undervoltage indication
The operation process is beeping (single measurement method)
Error message function
Manual and automatic shutdown functions
Meter Ketebalan Salutan TT270
TT270 thickness gauge head selection table is as follows:
| Probe model | F400 | F1 | F1/90° | F10 | |||
| Work Theory | Aruhan magnetik | ||||||
| Test range(um) | 0~400 | 0~1250 | 0~10000 | ||||
| Low resolution(um) | 0.1 | 0.1 | 10 | ||||
| Value error | One point calibration(um) | ±(3%H+1) | ±(3%H+10) | ||||
| Two point calibration(um) | ±[(1~3) %H+0.7] | ±[(1~3)%H+1] | ±[(1~3)%H+10] | ||||
|
Ujian syarat
|
Minimum radius of curvature(um) | convex | 1 | 1.5 | rata | 10 | |
| The diameter of the smallest area(mm) | 3 | 7 | 7 | 40 | |||
| Critical matrix thickness(mm) | 0.2 | 0.5 | 0.5 | 2 | |||
| Probe model | N1 | CN02 | |||||
| prinsip kerja | Eddy current | ||||||
| Measuring range(um) | 0~1250 | 10~200 | |||||
| Low resolution(um) | 0.1 | 1 | |||||
| Value error | One point calibration(um) | ±(3%H+1.5) | ±(3%H+1) | ||||
| Two point calibration(um) | ±[(1~3) %H+1.7] | — | |||||
| ujian
syarat
|
Minimum radius of curvature(um) | 3 | Only for flat | ||||
| The diameter of the smallest area(mm) | 5 | 7 | |||||
| Critical matrix thickness(mm) | 0.3 | unlimited | |||||