Automatic Dew Point Verification System Dew Point Calibration Test Machine
In this Dew Point Verification System, the humidity generator regulates humidity based on the principle of the divided flow method. By varying the mixing ratios of dry and humid gas streams, gases with different dew point temperatures can be generated. Depending on specific requirements, this method allows for multi-stage flow division configurations to produce a gas source with the desired dew point value. The system is capable of continuously generating humid gas with dew points ranging from -80°C to +20°C, with an adjustable flow rate of 0 kepada 10 L/min.
- Penerangan
- Siasatan
Penerangan
Automatic Dew Point Verification System Dew Point Calibration Test Machine
Gambaran keseluruhan:
In this Dew Point Verification System, the humidity generator regulates humidity based on the principle of the divided flow method. By varying the mixing ratios of dry and humid gas streams, gases with different dew point temperatures can be generated. Depending on specific requirements, this method allows for multi-stage flow division configurations to produce a gas source with the desired dew point value. The system is capable of continuously generating humid gas with dew points ranging from -80°C to +20°C, with an adjustable flow rate of 0 kepada 10 L/min.
Comprising units such as an air compressor, dry gas generator, dew point generator, and flow control valves, the system meets the operational requirements of several key standards: GB/T 5832.2-2016 (*Gas Analysis—Determination of Trace Moisture—Part 2: Dew Point Method*), JJG 499-2004 (*Verification Regulation of Precision Dew Point Meters*), and JJF 1272-2011 (*Calibration Specification for Capacitive Dew Point Meters*).
The entire system features a compact structure and a small footprint, while offering a high degree of automation. It is suitable for a wide range of industries, including metrology, higher education, scientific research, dew point meter manufacturing, kuasa elektrik, petrochemicals, gas asli, keluli, and microelectronics.
Key Features:
Integrated chassis design for convenient operation;
The gas source, dry gas generator, dew point generator, and dew point meter units can be independently detached and assembled, allowing for flexible configuration based on specific needs;
The total flow rate is adjustable from 0 kepada 10 L/min, allowing users to regulate the flow according to the number of devices under verification;
The dry gas generator is capable of stably outputting dry gas with a dew point temperature of < -80°C;
The dew point generator utilizes high-precision, high-stability mass flow controllers to mix the dry and humid gas streams, ensuring a stable gas output;
System operation is managed via a large 7-inch touchscreen display, providing a simple and user-friendly human-machine interface.
Automatic Dew Point Verification System Dew Point Calibration Test Machine
Equipment Requirements for the Dew Point Verification System:
Dry Gas Source (Dry Gas Generator)
Humidity Generator
Chilled Mirror Dew Point Meter
|
Spesifikasi Teknikal |
|
|
Sumber Gas |
Compatible with an air compressor, nitrogen cylinder, or liquid nitrogen cylinder; provides an operating pressure of 0.6–0.8 MPa; features simple maintenance and low operating costs. |
|
Dry Gas Generator |
Utilizes the proven Pressure Swing Adsorption (PSA) drying principle; the desiccant does not require replacement during long-term use.
Operasi mudah: simply adjust the pressure valve to control the output pressure. Real-time display of inlet and outlet pressures. |
|
Prinsip |
Desiccant Pressure Swing Adsorption |
|
Minimum Dry Gas Dew Point |
≤ -80°C |
|
Input Gas Pressure |
≥ 0.8 MPa |
|
Output Gas Pressure |
0.3 MPa (Boleh laras) |
|
Suhu Operasi |
5–40°C |
|
Suhu Penyimpanan |
-40–+50°C |
|
Bekalan Kuasa |
Ac 220 V ±10%, 50 Hz |
|
Penggunaan Kuasa |
300 W |
|
Dimensi |
485 mm × 500 mm × 266 mm |
|
Berat badan |
25 kg |
Humidity Generator
|
Spesifikasi Teknikal |
|
| Prinsip | Utilizes the shunt flow principle.
Features a two-stage pressure regulation design, with input pressure fluctuations of ≤0.005 MPa, ensuring system stability. Employs high-precision, high-stability mass flow controllers with rapid response times to ensure precise mixing ratios for dry and humid gases. |
| Julat | DP: -80°C to +15°C (Upper limit is 5°C below ambient temperature) |
| Kestabilan | DP: 30-minute fluctuation ≤ ±0.1°C |
| Masa Tindak Balas | ≤ 1 min |
| Hysteresis | None |
| Adjustable Flow Rate | 0–10 L/min |
| Keluaran | 3-way gas output: 1 output for connection to a standard reference, 2 outputs for working gas |
| Bekalan Kuasa | Ac 220 V ±10%, 50 Hz; Kuasa: 300 W |
| Dimensi / Berat badan | 485 mm × 500 mm × 266 mm; 18 kg |
| Suhu Operasi | 5–40°C |
| Suhu Penyimpanan | -40 to +50°C |