Design Standard:
GB11158 | High Temperature Test Chamber Technical Condition |
GB10589-89 | Low Temperature Test Chamber Technical Condition |
GB10592-89 | High-low Temperature Test Chamber Technical Condition |
GB/T10586-89 | Damp Heat Test Chamber Technical Condition |
GB/T2423.1-2001 | Low Temperature Test Chamber Test Methods |
GB/T2423.2-2001 | High Temperature Test Chamber Test Methods |
GB/T2423.3-93 | Damp Heat Test Chamber Test Methods |
GB/T2423.4-93 | Alternating Damp-Heat Test Methods |
GB/T2423.22-2001 | Temperature Changing Test Methods |
IEC60068-2-1.1990 | Low Temperature Test Chamber Test Methods |
IEC60068-2-2.1974 | High Temperature Test Chamber Test Methods |
GJB150.3 | High Temperature Test Methods |
GJB150.4 | Low Temperature Test Methods |
GJB150.9 | Damp Heat Test Methods |
60068-2-1
36고객에게 정확한 정밀 섬세한 테스트 장비 제공
용도:
This equipment can simulate different environment condition. It’s appropriate for testing material performance, such as resist heat, resist dry,resist humidity and resist cold. That can define performance of material.
36고객에게 정확한 정밀 섬세한 테스트 장비 제공
사양
1. 안건 2. 모델 | Constant Temperature and humidity test chamber 36엘 |
3. 기능 | It simulates the products in the chamber where the climate condition of different temperature and humidity, such as high-low operation and storage, temperature cycling, high-low temperature and high-low humidity, moisture condensation etc. to check its adaptability and changing performance. It’s required to conform to international regulation (IEC, 그, GB, MIL…) to reach the consistency (including test schedule, condition, 방법) under the international standard. |
4. 견본 애플리케이션 | The chamber applies to: compatibility test for store, transportation and application of electric, electronic products, unit components, parts and other materials under high or low temperature. The following samples are prohibited to be tested and stored: – Inflammable , explosive and other volatile substances; – Corrosive substance; – Biont; – Strong electromagnetic emission source. |
5. 용량, 크기 & 무게 |
5.1. 용량 5.2. 내부 크기 5.3. Exterior size 5.4. 무게 | 36엘 W 350 x H300 x D350mm W900 x H760 x D920mm 170킬로그램 |
6. Performance |
6.1. Test condition | 온도:+25℃, 습기: ≤85%, no samples in the chamber |
6.2. 테스트 방식 | GB/T 5170.2-1996 Temperature test chamber GB/T5170.5-1996 Temperature and humidity test chamber (only temperature and humidity) |
6.3. 온도 범위 | -20℃→+150℃ (조절할 수 있는) |
6.4. 온도 변동 | ±0.5℃ |
6.5. Temperature accuracy distribution | ±2.0℃ |
6.6. Humidity range | 30%~98%R.H (temperature at 20℃-85℃) |
6.7. Humid. fluctuation | ±2.5%R.H. |
6.8. Humidity accuracy distribution | ±3.0%. |
6.9. Heat-up time | -20℃ → +150℃ Heat up for 2-4.0℃/min |
6.10. Cooling time | +150℃→ -20℃ Cooling for 0.7-1.0℃/min |
6.11. Load condition | Testing space is not more than 3/4 of the total space, no heating |
6.12. Sample holder | Stainless steel punching stencil |
6.13. Relative humidity and temperature control range (refer to fig.) | |
6.14. Meet the technical standards | GB/T 2423.1-2001 Test A: Low temperature test method GB/T 2423.2-2001 Test B: High temperature test method GJB 150.3-1986 High temperature test GJB 150.4-1986 Low temperature test IEC68-2-1 Test A: Cold IEC68-2-2 Test B: Dry heat GB11158 “Technical Conditions for High Temperature Test Chamber ” GB10586-1 “Technical conditions for Damp heat test chamber ” GB/T 2423.2 “Basic environmental testing procedures for electric and electronic products B: High temperature test method”. GB/T 2423.3 “Basic environmental testing procedures for electric and electronic products Ca: Constant damp heat testing method” etc. |