コーティング厚さ計のトラブルシューティング方法

非破壊検査における膜厚計の主な欠点は、数値表示が安定しないこと, 大きな誤差, 数値表示なし, 等.The causes of these failures are from the instrument itself, 測定したワークから, 自然環境からも. これらの障害をトラブルシューティングする方法は次のとおりです。.

コーティング厚さ計のトラブルシューティング方法

1. Display value is not stable

The cause of the instability of the coating thickness gauge is mainly due to the particularity of the material and structure of the workpiece itself, such as whether the workpiece itself is a magnetic conductive material. If it is a magnetic conductive material, we have to choose the magnetic coating thickness gauge. If the workpiece is conductive, we have to choose the eddy current coating thickness gauge.There are workpiece surface roughness and attachment is also the cause of the instrument value display temperature, workpiece surface roughness is too large, surface attachment too much.The key to troubleshooting is to grind the workpiece with large roughness, go out to the attachment, and then choose the appropriate coating thickness tester.

2. There are too many errors in measurement results

The reasons for the large measurement error of the coating thickness gauge have been introduced clearly in the previous articles. Here we briefly introduce the main reasons for the large measurement error:Matrix metal magnetization, matrix metal thickness is too small, edge effect, workpiece curvature is too small, surface roughness is too large, magnetic field interference probe placement method, new friends can refer to the instrument article familiar with we will not do a introduction.

3. Don’t display numbers

The simple reason why the coating thickness gauge does not display a number is to check whether the battery is sufficiently charged. If the battery is sufficiently charged, the measurement still does not display the value.