Ozone aging test chamber
io. Applicazioni
Ozone, though present in very low atmospheric concentrations, is a major factor in rubber cracking. This product simulates and intensifies atmospheric ozone conditions to study the effects of ozone on rubber and provide a method for rapidly identifying and evaluating rubber’s resistance to ozone aging and the protective effectiveness of antiozonants.
Reference standards include: GB/T7762-2003, Vulcanized or Thermoplastic Rubber – Static Tensile Test Method for Ozone Cracking Resistance, GB/T2951.21-2008, General Test Methods for Cable Insulation and Sheathing Materials, GB / T 11206-2009, Rubber Aging Test – Surface Cracking Method, and test methods specified in other relevant standards.
II. Principle and Method
Test methods can be categorized as “statico” o “dinamico,” depending on the test method and standard. Static testing involves stretching the sample before placing it in a test chamber for testing. Dynamic testing involves placing the sample in the test chamber’s fixture and testing while it is stretched. The tensile fluctuation should be approximately 5%-45% of the sample itself.
III. Structural Features
3.1. Struttura: All-in-one unit.
3.2. Materiali
Inner Box: Acciaio inossidabile SUS304
Outer Box: SUS304 stainless steel or painted
Insulation: Dual protection with rigid polyurethane foam and aluminum silicate insulation.
3.3. Door – Single door, opening on the left.
Two silicone rubber seals on the window and door frames, and an electric heater to prevent condensation.
Observation windows (W300 x H450 mm) and lighting are provided above the doors and windows.
3.4. Sample Holder Requirements
un) (Dynamic Test; preparazione regolabile) – Secure the specimen at both ends with clamps at the required elongation. When exposed to ozone-containing air, the specimen’s length should be approximately parallel to the airflow.
B) The clamps should be made of a material that does not readily decompose ozone (per esempio., alluminio). C) Maintain a specimen rotation speed between (20 e 25) ~199,9 mm/s (valore effettivo). Each specimen should continuously move along the same path within a plane perpendicular to the dry airflow.
D) The time required for one rotation of the specimen is (8 a 12) minuti. The specimen scanning area should be at least 40% of the effective area of the test chamber.
ed075fd1b9d3) Specimen preparation should comply with the requirements of GB/T 9865.1. Specimens should preferably be cut from freshly molded test pieces; se necessario, they may be cut from finished products. A minimum of three specimens should be used for each test condition. Standard strip specimens should be at least 10 mm wide and 2.0 mm± 0.2 mm di spessore. The specimen length between the two ends of the grips before tensioning should be at least 40 mm.
3.5. Control Panel – Controller Display
Power Switch
Light Button
3.6. Chamber Configuration – One internal stainless steel SUS #304 rack turntable
One φ50 mm wire hole, with a cover and silicone rubber plug.
4 movable casters
4 horizontal fixed legs
A drain hole is located at the lower rear of the unit to facilitate condensate drainage.
Ozone aging test chamber
Parametri tecnici
| Intervallo di temperatura | 0°C-70°C |
| Fluttuazione della temperatura | ±2.0°C |
| Ozone Concentration Accuracy | 10-1000 pphm ±10% |
| Sample Holder Lifting Frequency | 2 R/min |
| Airflow Speed | 12-16 ~199,9 mm/s (valore effettivo) |
| Test Gas Flow Rate | 20-70 l/min |
| Sample Holder Rotation Speed | 360° (1-3 giri/min) |
| Humidity Range | 20-98% |