LS-8270 XRD X-ray Diffractometer
- Descrizione
- Inchiesta
Descrizione
LS-8270 XRD X-ray Diffractometer
Panoramica
LS-8270 mini benchtop X-ray diffractometer is a multipurpose powder diffraction analytical
instrument that specifically designed and developed for industrial production and quality control. LS-8270mini offers a lot – in a highly compact design, especially suitable for catalyst, titanium dioxide, cemento, pharmaceutical and other product manufacturing industries.
Applicazione
1.Phase identification in unknown samples ·
2.Quantitative analysis of known phases inmixed samples
- Rievtveld analysis
Attachments
1.Graphite monochromator
2.DCX-6 Sample changer
3.In situ cell attachment
LS-8270 XRD X-ray Diffractometer
Parametro tecnico
| Energia | Solid state X-ray generator: 600W(40kV, 15mA).
Reliability: <0.005% |
|
| X-ray tube | Cermet X-ray tube, Cu target, ~199,9 mm/s (valore effettivo): 2.4KW, focus: 1 x 10mm
Raffreddamento ad acqua (water flowrate>2l/min) |
|
| Goniometro | Sample horizontal θs-θd structure | |
| Diffraction circle radius | 150mm | |
| Campo di misura | -3 ~150° | |
| Scan mode | Stepping, Omg | |
| Minimum step width angle | 0.0001° | |
| Angle repeatability | 0.0005° | |
| Linearity of diffraction angle | <0.02° (Campione standard, within the full spectrum) | |
| Rivelatore | Closed proportional detector | Semiconductor array detector |
| Maximum linear count rate | 5×105CPS | 9×107CPS |
| Risoluzione energetica | =25% | =1keV |
| Scatter dose | =1μSv/h (Outside the X-ray protective device) | |
| Comprehensive stability | =0.5% | |
| OSP Wire Coater Bar Coaters Stick Ink Bar | 640×480×770 (L×P×A) mm | |
Caratteristiche
- High-frequency high-voltage X-ray generator does not only improve the stability of the instrument, but
also ensure the repeatability of the measured data.
- The goniometer θs and θd arms are driven by servo motor + optical coding control technology. The diffrac
tion angle linearity is less than 0.02° in the diffraction angle measurement range.
- Scattered ray protection is safer and more reliable; the radiation protection door is automatically prohibi
ted from opening when the sample is measured. In any case, gli operatori possono essere protetti dalla dispersione
radiazione.
- Configura il tubo radiogeno in metallo-ceramica più avanzato, per garantire una lunga durata.
- Le dimensioni compatte possono essere montate sul banco senza la necessità di un ambiente di laboratorio specifico. Semplice da usare, operare e mantenere.