- Descrizione
- Inchiesta
LS-201-4 2 Camera di umidità della temperatura della porta
1.Specifica
Modello | LS-201-4-80D |
Up inner size | W400*H500*D400mm |
Down inner size | W400*H500*D400mm |
Dimensioni esterne | W1000*H1720*D1500mm |
Intervallo di temperatura | 0℃~150℃ |
Intervallo di umidità | 20%~98%RH. |
Inner material | SUS304 stainless steel |
Out material | SUS304 Stainless steel or electrostatic spray painting |
Cool system | French TECUMSEH double stage compressor |
refrigerating fluid | R404a and R23(DUPONT) |
Cooling way | Air cooled |
Controllore | Independently developed 7 inch full-color touch screen controller |
Heating rate | 2.0℃~4.0℃/min nonlinear |
Cooling rate | 0.7℃~1.0℃/min nonlinear |
Accessories | 50mm connection hole *1 impostato, wet gauze *1 impostato, 4GB usb disk (with operation video)*1 impostato, manual warranty card *1 impostato, and rack *2 pezzi |
Peso(riflettività di scissione della luce) | 500 |
Energia | 380V3phase |
Rate of (kw) | 6 |
Current(UN) | 15 |
LS-201-4 2 Camera di umidità della temperatura della porta
2.Specifica tecnica
1. Equipment use | Equipment for testing the properties of materials in various environments and testing the heat, cold, dry and moisture resistance of various materials.It is suitable for testing quality of electronic, electrical, mobile phone, communication, instrument, vehicle, prodotti in plastica, metallo, cibo, chemistry, materiali da costruzione, medical and aerospace products. |
2. Test sample | This test equipment is prohibited: Test or storage of samples of inflammable, explosive and volatile substances or of samples of stored corrosive substances or of stored biological samples or of samples of strong electromagnetic emission sources. |
3. Prestazione | |
3.1. Test environmental conditions 3.2. Metodo di prova | Environment temperature is 25 ℃, relative humidity of 85% or less, test under the condition of no sample in the box GB / T 5170.2 Temperature testing equipment GB / T 5170.5 Wet and hot test equipment (wet and hot only) |
3.3. Temperature fluctuations | ±0,5℃ |
3.4. Temperature deviation | ±2,0 ℃ |
3.5. Humidity fluctuations | ±2,5%UR. |
3.6. Humidity deviation | ±3,0%. |
3.7. load | The test space is no more than three quarters of the total space, no heating source。 |