Dynamischer thermomechanischer Analysator
Dynamic Thermomechanical Analyzer applies single-frequency or multi-frequency oscillatory forces to samples at a programmable temperature to study their mechanical behavior. The instrument measures deformation, Speichermodul, loss modulus, and loss factor as a function of temperature, Zeit, and force frequency. Suitable specimens include thermoplastics and thermosetting plastics, Gummi, Beschichtungen, metals and alloys, anorganische Materialien, und Verbundwerkstoffe.
- Beschreibung
- Anfrage
Beschreibung
Dynamischer thermomechanischer Analysator
Key Performance Indicators
The instrument applies single-frequency or multi-frequency oscillatory forces to samples at a programmable temperature to study their mechanical behavior. The instrument measures deformation, Speichermodul, loss modulus, and loss factor as a function of temperature, Zeit, and force frequency. Suitable specimens include thermoplastics and thermosetting Kunststoffe, Gummi, Beschichtungen, metals and alloys, anorganische Materialien, und Verbundwerkstoffe.
Dynamischer thermomechanischer Analysator
Technische Parameter
| Temperaturbereich | -150°C to 600°C |
| Instrument Structure | Circular, sealed furnace ensures uniform heating, with bottom-mounted sample loading. |
| Heizrate | 0.1~ 20°C/min |
| Temperaturgenauigkeit | ±0,1°C |
| Frequenzbereich | 0.01~100 Hz (Selectable) |
| Force Range
|
1mN~18N; Minimum 0.001N (dynamisch)/0.005N (statisch), Maximum 50N, Resolution 0.0001N |
| Verschiebungsauflösung | 1nm |
| Deformation Range | Maximum 25mm |
| Amplitude | ±0.5~10000µm |
| Modulus Range | 10³~10¹² Pa |
| Modulus Accuracy | 0.5% |
| Damping Range (tanδ): | 0.005~ 100 |
| Tanδ Resolution | 0.00005 |
|
Sample Temperature Control (STC) Funktionen, software screenshots provided. |
|
|
Deformation Modes: Compression, Penetration (maximum testable sample diameter no less than 15mm, maximum thickness no less than 25mm); Zugfest (maximum width no less than 13mm, maximum thickness no less than 5mm, maximum length no less than 30mm); Scheren, Three-Point Bending, Single/Dual Cantilever. |
|
|
Automatically identifies installed sample holders, automatically calibrates, and automatically measures sample length. |
|
|
The software features test editing, data acquisition, Lager, Verarbeitung, and analysis functions; software controls purge and shielding gases. 1.Software Features Experimental Design Curing Simulation: Supports step-by-step temperature ramping and constant temperature testing, automatically recording modulus-time curves. Zeit-Temperatur-Äquivalenz (TTS): Erstellt Masterkurven für langfristige Leistungsvorhersagen. 2. Softwarefunktionen Datenverarbeitung Automatische Berechnung: Echtzeitausgabe des Speichermoduls, Verlustfaktor (tanδ), und Glasübergangstemperatur (Tg). Multi-Graph-Vergleich: Unterstützt die Overlay-Analyse von Daten bei verschiedenen Frequenzen und Temperaturen. |
|
|
Optionales Sonderzubehör |
Feuchtigkeitszubehör |
| Optionale Vorrichtungen | Schervorrichtung, Doppelauslegerbefestigung, Flüssiges Gerät, Dreipunkt-Biegevorrichtung |